B46600E-M

600W Open Frame Medial Power Supplies

This series of 600W Open Frame Medical Power Supplies are designed for medical applications.

Universal AC Input of 90 – 264VAC, 47 – 63Hz and a typical efficiency of >90% with no load energy consumption of <500mW (for 5Vsb standby mode) and active PFC >0.90 @ 230VAC and full load.

Output range of 12 – 48VDC and 0A minimum load which features low ripple and noise (300mV p-p Max.).  The range also includes in-built protection for over temperature overload, OVP and SCP. output regulation is +3%/-3% Max.

This series is designed for both medical and ITE applications and meets medical safety to EN60601-1 and general safety to EN62368-1 with UL, cUL, TUV and CB approvals. It complies with EMC standards EN60601-1-2, EN55011, EN55032 Class B, EN55035, FCC Part 15 B and FCC Part 18 B.

Unit dimensions for this product range is 152.4 x 101.6 x 42.0mm and achieves full output power with convection or force air cooling options. The series also features terminal block input and output connectors. The operating temperature range is -20 to +70°C (refer to derating curves).

This series is designed to cover a wide range of modern applications. These include, but are not limited to office equipment, IoT equipment, analytical equipment, industrial machines and consumer products for both medical and ITE applications. 3 year warranty.

Features

  • Both ITE & Medical Approvals.
  • Medical applications Protection: Means of Patient Protection ( MOPP).
  • Meet Medical BF rated.
  • High Power density, 600W in 6″ x 4″ footprint.
  • Altitude during operation: ITE Below 5000m. & Medical Below 3000m.
  • No load Power consumption < 500mW (for 5Vsb standby mode)
  • Meet RoHS requirement
  • 3 year warranty
  • Input range : 90 – 264VAC (Refer to derating curve)
  • Frequency : 47 – 63Hz. RoHS compliant
  • Power Factor : > 0.95 @115VAC; > 0.90 @230VAC @full load.
  • Inrush current :<40A peak @115VAC; <80A peak @230VAC cold start @25°C. U-Channel type
  • Input current (rms) : 8A @115VAC; 4A @230VAC max.
  • Efficiency : > 93% typical @half load, 230VAC.
  • Touch current < 100uA @264VAC.
  • Maximum output power : 600Watts forced air, 300/340Watts convection cooling.
  • Hold-up time : > 10ms typical @full load, 115VAC.
  • Short circuit protection : Auto-recovery.
  • Over power protection : 105% to 130% maximum rating, Auto-recovery.
  • Over voltage protection : Latching type. AC Recycle or SD pin is reset.
  • Over temperature protection : Latching type. AC Recycle or SD pin is reset.
  • Standby and FAN output supply
  • Remote control (Inhibit) function

Products in the Series

SeriesPart NumberOutput Power (W)Input VoltageOutput Voltage (VDC)Output Current (A)
B46600-12E-M60090 - 264VAC1250
B46600-24E-M60090 - 264VAC2425
B46600-28E-M60090 - 264VAC2821.43
B46600-36E-M60090 - 264VAC3616.67
B46600-48E-M60090 - 264VAC4812.5